Hangzhou Dongzi Import & Export Co., Ltd
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Grid-control,70Kv 8mA Diagnostic Dental X-ray Tube for Intra-oral X-ray machines
Quick Detail:
Type:Station anode x-ray tube
Application: For intra-oral dental x-ray unit
Model: KL2-0.8-70G
Equivalent to CEI OCX/65-G
Integrated high quality glass tube
Description:
KL2-0.8-70G Stationary anode X-Ray Tube is especially designed for intra-oral dental x-ray unit and available for a nominal tube voltage with self-rectified circuit. And it is a grid-control tube.
The integrated high quality tube with glass design has one super imposed focal spot and a
reinforced anode. It is extremely important to observe the connection diagram and the grid-resistor value. Any change could modify the dimensions of the focal spot, also varying diagnostic performances or overloading anode target.
The high anode heat storage capacity ensures a wide range of applications for intra-oral dental application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.
Applications:
KL2-0.8-70G Stationary anode X-Ray Tube is especially designed for intra-oral dental x-ray unit and available for a nominal tube voltage with self-rectified circuit.
Specifications:
Nominal Tube Voltage……………………………………..….………..…….…………..…70Kv
Nominal Inverse Voltage…………………………………..…..…….…..………………….85kV
Nominal Tube Current………..……………………..….……….………………………...…8mA
Maximum Exposure Times………………………………………….……….……........ …...3.2s
Maximun anode cooling Rate………………………………………………………………210W
Max. Anode Heat Content………….……….…………………..……………………..…...7.5kJ
Filament Characteristics…………………………………………..Uf=4.0V(Fixed),If=2.8±0.3A
Focal spot ..…………………………………………………….……….….0.8(IEC 60336 2005)
At 70kV 8mA with 5kΩ to 25 kΩbias resister(Fixed)
Grid Resistance Value…………………………recommended by manufacturer for any tube
Target Angle……………………………………………………………….………………..…19°
Target Material………………………………………………………………….………..Tungsten
Cathode type……………………….…………………………………………….…….W filament
Permanent Filtration………..…………..……………...Min. 0.5mmAl/50 kV(IEC60522/1999)
Dimensions………………………………..…………….……80mm length by 30mm diameter
Weight…………………………………………………………………………….…….125 grams
Competitive Advantage:
Elevated anode heat storage capacity and cooling
Constant high dose yield
Excellent life time